Comparative Study of Conventional Drill Bits and a New Model for Low-rotation in the Surgical Bed Preparation in Bone Blocks for Installation of Dental Implant

Many works in Dentistry have aimed toward improving bone regeneration and oral rehabilitation through techniques, procedures and materials of Implantology. Objective of this study is to comparethe heating intensity of bovine blocks, simulating surgical beds using different drills.

Kopp Implantes
by Kopp Implantes
Comparative Study of Conventional Drill Bits and a New Model for Low-rotation in the Surgical Bed Preparation in Bone Blocks for Installation of Dental Implant

Primo Herrera Subelza1,*, Gino Kopp², Maurício Fernando Herrera Sivila³, Henry Kevin Herrera Sivila4, Carlos Eduardo Francischone1,5
1São Leopoldo Mandic School of Dentistry, Campinas, São Paulo, BRAZIL.
2Instituto Kopp Odontologia, Curitiba, Paraná, BRAZIL.
3Centro Odontologico Herrera, Sucre, BOLIVIA.
4Universidad Mayor Realy Pontificia San Francisco Xavier de Chuquisaca, Sucre, BOLIVIA.
5University of São Paulo, Bauru, São Paulo, BRAZIL.

ABSTRACT
Background: Many works in Dentistry have aimed toward improving bone regeneration and oral rehabilitation through techniques, procedures and materials of Implantology. Objective of this study is to comparethe heating intensity of bovine blocks, simulating surgical beds using different drills.
Materials and Methods: We evaluated 24 blocks divided into 4 groups with different KOPP brand drill models: G1-Conventional drills without irrigation; G2-Experimental drills without irrigation; G3-Conventional drills with irrigation; G4-Experimental drills with irrigation. Perforations were performed with depths of 1.0 and 5.00 mm, drill rotation speed of 150 rpm
and torque of 45 N. Two K-type thermocouples were used to measure the temperature of the bone, with the first in the 1 mm deep perforation andthe second in the 5 mm perforation. Results: In comparing the conventional and experimental drills treated with the same kind of protocol there was no statistical difference in the depth of 1 mm (p> 0.05), however a statistical difference was observed at the depth of 5 mm (p< 0.05). When comparing the temperature of each drill there was statistical difference between the groups treated with no irrigation (G1 and G2) compared to the groups treated with the irrigation protocol (G3 and G4) and the last two had a significant reduction in the temperature of the two studied depths (p<0.001). Conclusion: The groups of the bovine blocks treated with irrigation showed lower temperature records when compared to the groups without irrigation protocol in both studied depths.
Key words: Bone, Dental implants, Irrigation, Osseointegration, Surgical cutters.
Correspondence
Dr Primo Herrera Subelza,
School of Dentistry, 13 José Rocha Junqueira, 13045-755- Pte. Preta, Campinas, São Paulo, BRAZIL.
Phone no: +55 1932113600
Email: [email protected]
DOI: 10.5530/jyp.2019.11.88

INTRODUCTION
Edentulism is a condition that exists in all countries, however the technological advancement of Dentistry has enabled the development and emergence of many means of oral rehabilitation.1 With the clinical and scientific proof of osseointegration, dental implant treatment has been consolidated as a reality in the dental clinic. Thus, the search for osseointegration and preservation of peri-implant bone tissue has been intense in recent years.2,3 Many works in Dentistry in recent years have aimed toward improving bone regeneration and oral rehabilitation through techniques, procedures and materials placed on the market of Implantology.

Industries are increasingly investing in innovation of products to ensure competitiveness and to remain on the market. A properly planned surgery with theproper choice of the implant type, surgical technique and appropriate biosecurity will likely achieve clinical success.4 Nowadays, various types and formats of implants are constantly released on the market. The development of new materials is due to the search for improving implants to prevent failures on osseointegration.5

There are several variables in implant failure prevention, such as surface modification, implant material, implant geometry, training and professional technique. All these factors may decisively influence osseointegration. 5 When the surgical bed preparation for installing dental implants with conventional surgical cutters is completed, the aggression of osteogenic cells, blood cells and undifferentiated cells is inevitable due to friction by bone compression which ends up causing osteonecrosis and cell death of important elements that may jeopardize the osseointegration.6,7

According to Nero et al.8 the temperature threshold that the bone tissue can reach is between 44°C to 47°C for one min. Some of the factors which influence increasing this temperature is the rotational speed and the drill drawing. A protocol held at low speed decreases the stress caused to the bone and the bone resorption is smaller. Therefore, the objective of this study was to evaluate different irrigation techniques in controlling boné temperature during osteotomies, comparing a conventional system to na experimental one for the installation of dental implants (INPI, BR 20 2017 010735 9).9

MATERIALS AND METHODS
Group Division

A methodology similar to that of Nero et al.8 was followed in this study. The motor and contra-angles used for the perforations were both of the Driller brand, Brazilian industry. The surgical contra-angle was fixed to the equipment, always allowing the same angulation and the same reduction in all osteotomies. This experimental model shows a lower area of contact and consequently lower friction during the making of the surgical alveolus. Also has better cutting power at the end due to front and side edging. The macro structure of the drill has a chamfer of 180 degrees and undulations in the longitudinal direction (Figure 1).

The surgical motor was coupled to the contra-angle guaranteeing the same rotational speed (150 rpm) of the drills and the same torque. A total of 24 bovine blocks were donated, approximately 14 cm long 13mm wide, 20 mm high and approximately 2 mm cortical bone (Figure 2).

The blocks were previously prepared and divided into four groups. We evaluated 24 bovine blocks divided into 4 groups, 6 blocks per group for analysis in triplicate, with different KOPP brand drill models (Brazilian industry), evaluated under rotation of 150 rpm: G1 – conventional drills without irrigation; G2 – experimental drills without irrigation; G3 -conventional drills with irrigation; G4 – experimental drills with irrigation. The perforations were performed with depths of 1.0 and 5.0 mm, rotational speed of the drill at 150 rpm and torque of 45 N.

Two K-type thermocouples attached to a portable digital thermometer (Tasi, model 8620) were used to measure the bone temperature and the thermometer was calibrated to measure the temperature at the depths of 1mm and 5mm. The edges of the thermocouples were stabilized with photopolymerizable resin. A thermostatic box was used so that the experiments were carried out at a temperature of 37°C and the blocks were 3/4 covered, thus simulating a moisture approximate to that of the oral cavity (Figure 3).

The data were tabulated in Microsoft Office Excel® software and exported to SigmaPlot version 11.0 software. The analyses were performed adopting a confidence level of 95% (P < 0.05). The data were expressed as means and standard error. The quantitative data were analyzed by the normality test and compared using the Student’s t-test and Chi-squared test, as appropriate.

RESULTS
When evaluating the mean temperature values at 1 mm depth in the
different protocols (with and without irrigation), the mean temperature value in G1 was 25.95 ± 0.46°C. Maintaining the same previous
characteristics except only changing the experimental drills (G2), the
mean value was 25.22 ± 0.44°C. Furthermore, the mean value in G3 was
21.01 ± 0.52°C and in G4 the mean value was 19.79 ± 0.48°C. When
evaluating the mean temperature values at 5 mm depth in the different
protocols (without and with irrigation), the mean temperature value in
G1 was 30.59 ± 0.47°C. Maintaining the same previous characteristics
except only changing the experimental drills (G2), the mean value was
29.74 ± 0.41°C. In G3 the mean value was 27.27 ± 0.39°C and in G4
the mean value was 24.91 ± 0.42°C. There was no statistical difference
between the groups.

Figure 1: Conventional and experimental drill designs, respectively.
Left: Conventional drills; Right: Experimental drills.
Figure 2: Adaptation of bovine bone and marking of points to be pierced.
Preparation of the model to simulate the surgical bed.
Figure 3: Calibration of the depths and adaptation of thermocouples;
Sample immersed in thermostatic water box and temperature record
(below). Thermocouples and preparation of the bones.
Figure 4: Average values of the block temperatures at depths of 1 mm and
5mm at 150 RPM with and without irrigation with conventional drills and
experimental KOPP brand drills.

Table 1 shows the intra and inter comparison of the block temperatures
without irrigation with conventional drills. The minimum temperature
at the depth of 1 mm occurred in the 3.0 mm drill = 24°C and the
maximum in the 3.8 mm drill = 27.5°C, while the minimum temperature
at the depth of 5 mm was 28.7°C in the 2.0 mm drill and 32.1°C in the
4.8 mm drill. There was statistical difference between the temperatures in the intragroup comparison of each depth (p< 0.001), however there was
no statistical difference in the Intergroup analysis (p= 0.22).

Table 2 shows the different values in the experimental protocol without
irrigation. The minimum temperature at the depth of 1 mm occurred in
the 3.0 mm drill= 23.3°C and the maximum at the spear= 27°C and the
minimum temperature at the depth of 5 mm was 28.0°C in the 2.0 mm
drill and 31.1°C in the 4.8 mm drill. There was a statistical difference
between the temperatures in the intragroup comparison of each depth
(p< 0.001), while there was no statistical difference in the Intergroup
analysis (P = 0.25).

Table 3 describes the temperatures in the irrigation protocol with
conventional drills. The minimum temperature at the depth of 1 mm
occurred in the 2.0 mm drill= 18.5°C and the maximum in the 2.8-4.8 mm
drill= 21.8°C, while the minimum temperature at the depth of 5 mm was
26.2°C in the 2.0 mm drill and 28.9°C in the drill 2.8 mm. (Table 3). The
minimum temperature at the depth of 1 mm occurred in the 2.0 mm
drill= 18.5°C and the maximum in the 2.8-4.8 mm drill= 21.8°C, while
the minimum temperature at the depth of 5 mm was 26.2°C in the 2.0 mm
drill and 28.9°C in the 2.8 mm drill (Table 3). There was a statistical
difference between the temperatures in the intragroup comparison of
each depth (p< 0.001), but there was no statistical difference in the Intergroup analysis (P= 0.78).

Table 4 describes the values of the experimental protocol with irrigation.
The minimum temperature at the depth of 1 mm occurred in the spear
drill= 18°C and the maximum in the 4.3 mm drill= 22.1°C, while the minimum temperature at the depth of 5 mm was 22.3°C at the spear and
26.1°C in the 2.8 mm drill. There was a statistical difference between the
temperatures in the intragroup comparison of each depth (p< 0.001), but
there was no statistical difference in the Intergroup analysis (P= 0.26).

Table 5 compared the different temperatures in the protocols with and
without irrigation using each drill at the depth of 1mm. No statistical
difference was observed when comparing conventional and experimental
drills treated with the same protocol (p> 0.05). There was statistical
difference (p < 0.001) when comparing the temperature of each drill
between the groups treated without the irrigation protocol (G1 and G2)
when compared to the groups treated with the irrigation protocol
(G3 and G4). The last two had significant reduction in the temperature
record of all drills.

Table 6 compares the different temperatures in the protocols with and
without irrigation using each drill at the 5mm depth. There was a statistical
difference (P< 0.05) when comparing conventional and experimental
drills treated with the same protocol. There was also statistical difference
(p< 0.001) when comparing the temperature of each drill between the
groups treated without irrigation protocol (G1 and G2) when compared
to the groups treated with irrigation protocol (G3 and G4). The last two
had significant reduction in the temperature record of all drills.

Table 1: Description of the block temperatures at each depth at 150 RPM without irrigation with conventional KOPP brand drills.

mm= millimeters; values in degree Celsius (°C), * P <0.001; In-group Student’s
t-test, ** Chi-square test between 1 mm vs. 5 mm depths.

Table 2: Description of block temperatures at each depth with 150 RPM
without irrigation with experimental KOPP brand drills.

mm = millimeters; values in degree Celsius (°C), * P <0.001; intragroup Student’st-test, ** chi-squared test between 1 mm vs. 5 mm depths.

Table 3: Description of the block temperatures at each depth at 150 RPM
with irrigation using conventional KOPP brand drills

mm = millimeters; values in degree Celsius (°C), * P< 0.001; intragroup Student’s t-test, ** chi-squared test between 1 mm vs. 5 mm depths.

Table 4: Description of the block temperatures at each depth at 150 RPM with irrigation and experimental KOPP brand drills.

mm= millimeters; values in degree Celsius (°C), * P<0.001; intragroup Student’st-test, ** chi-squared test between 1 mm vs. 5 mm depths.

Table 5: Comparative analysis of the block temperatures at the depth
of 1 mm at 150 RPM with and without irrigation with conventional and
experimental KOPP brand drills.

mm = millimeters; values in degrees Celsius (°C), * ANOVA followed by the Holm-Sidak test when comparing intergroups; ** P> 0.05; ANOVA followed by the Holm-Sidak test in the intragroup comparison.

Table 6: Comparative analysis of the block temperatures at the depth of
5 mm at 150 RPM with and without irrigation using conventional and
experimental KOPP brand drills.

mm= millimeters; values in degrees Celsius (°C), * ANOVA followed by the Holm-Sidak test when comparing intergroups; ** P> 0.05; ANOVA followed by the Holm-Sidak test in the intragroup comparison.

DISCUSSION
The success of rehabilitation with implants in clinical practice is undeniable, however failures of multifactorial origin which are often combined can occur, thereby causing damage to the professional and the patient.

Among these failures, bone overheating stands out as well as poor case
planning. This elevation in temperature occurs due to the friction of the
surgical drill with the bone tissue and can cause damage to osseointegration and even necrosis when it crosses the limit between 44°C and 47°C
for at least one min.¹0 Devices with thermocouple technology can be used to check the bone temperature after drilling.¹¹ Thermocouples are electrical devices with great application in temperature measurement. In the study by Nero et al.8 they were used to evaluate the temperatures in the medullary bone and cortical bovine during osteotomy with a 2.00 mm drill. External irrigation was used in the halves of the specimens, while double irrigation was used in the remaining which was more efficient in temperature control. In the present study, the groups treated with irrigation (G3 and G4) showed better temperature control when evaluating the mean values at the perforation temperatures of 1.00 mm depth (G1: 25.95 ± 0.46°C; G2: 25.22 ± 0.44°C; G3: 21.01 ± 0.52°C; G4: 19.79 ± 0.48°C) and 5.00mm
depth (G1: 30.59 ± 0.47°C; G2: 29.74 ± 0.41°C; G3: 27.27 ± 0.39°C; G4:
24.91 ± 0.42°C). However, there were no statistical differences between
the different groups.

Andriani Jr. developed a thermocouple technology device associated
with a computerized system to evaluate the performance of irrigation
systems (internal and external) during the preparation of the surgical
bed for implant placement. The author evaluated 20 specimens of
bovine bone associated with thermocouples placed at different depths
and observed that the internal irrigation system generated lower heat
when compared to the external irrigation system. In the present study,
bovine bone was also used for presenting characteristics such as density
and similar configuration to the human jaw.¹¹

Carneiro et al.¹² evaluated the influence of the cut conditions on bovine
bone, monitoring the temperature through thermocouples. The authors
utilized a kit of uncoated dental drills using water for irrigation. In this
study, helical drills produced lower temperature values. It is known that
the vascularization around the implant is one of the main responsible
factors for the success of osseointegration. Maintaining this tissue during
the milling process is essential. Histological studies indicate that
temperatures higher than 47°C can inhibit bone regeneration.

Faria¹³ evaluated the temperature generated during the preparation of the
bovine bone tissue by comparing the Nobel and 3i connection systems.
After perforations with drills of 2.0mm and 3.0mm diameter up to
13mm depth, at a speed of 1500 rpm with intermittent pressure of 2kg
under constant irrigation at room temperature (24±1°C), it was
observed that the 2mm diameter connection system had better results
in relation to minimizing the temperature increase; however, none of
the drills exceeded the biological limit. Factors such as motor rotation
speed, drill design and pressure should be controlled. In the intragroup
analysis of each studied protocol herein, there was a statistical difference
between the temperatures of the different drills at each depth (p< 0.001).
However, there was no statistical difference in comparing the values of
the different depths studied in the Intergroup analysis.

Geerts and Patel6 evaluated the temperature changes along the surface
of the implant in vitro through thermocouples at ten min intervals and
established that the temperature threshold that the abutment could reach
is 47°C/1 min. In a total of 53 series of tests, the temperature of the
abutment varied between 52.80°C and 71.72°C, having an association
between the maximum temperature, the implant level and the temperature
of the abutment. The threshold of 47°C/1min was reached 31 times at the
implant level and the temperature was lower in areas farther away from
the heat source. For an abutment temperature of 62.3°C there was a 50%
chance that 47°C would be reached for 1 min at the implant level.
Gehrke et al.14 evaluated whether there were differences in temperature
variation using only a single drill for the preparation of surgical drill and
compared with the conventional drill sequence using various drills in
bovine bone. Three groups were evaluated: one group using three
consecutive 4.1mm cylindrical drills, one using three consecutive drills
for a 4.3mm conical implant and a third group with only one drill for a
4.2mm conical implant. The temperature in the cortical bone was
measured through a thermocouple. The same experiment with external
irrigation was repeated without irrigation, in which the protocol with
only one drill showed no greater bone heating than the conventional one.
When comparing temperature values when using each drill in the different
protocols in the present study, the treated groups without irrigation had
a significantly higher temperature than those with irrigation protocol
(p< 0.001).

Soldatos et al.15 measured the temperature in the surgery of different
implant designs, including a conical, a self-tapping conical and a parallel
wall design. The drills were compared with and without external irrigation
at rotations of 800, 1000 and 1200 rpm. The highest temperature was
found with the initial drills for conical implants and the lowest for the
initial drills for self-tapping conical implants. Irrigation especially
influenced the self-tapping design. The authors concluded that different
implant designs have different thermodynamic behaviors.

Strbac et al.16 evaluated changes in temperature during osteotomy in
bone samples. For this purpose, 160 automatic intermittent osteotomies
(10/16 mm deep drilling) were performed with 2 mm helical drills and
3.5 mm conical drills with and without irrigation. In their study, it was observed that the greatest changes in temperature were during the drill removal with the influence of irrigation. According to the authors, the progressive increase in temperature may modify the mineral structure of the bone hydroxyapatite.

Several studies have shown that atraumatic surgical techniques that prevent the implant site overheating and primary stability are the two main factors involved in the success of osseointegration.17 Bone perforation is a highly contested topic due to the heat production caused by this procedure. Several factors interfere with bone drilling and it is essential to develop new technologies in order to control them.

CONCLUSION
In the present study it can be concluded that the groups of bovine blocks
treated with irrigation had lower temperature records when compared to
the groups without irrigation protocol for both studied depths.
ACKNOWLEDGEMENT
We acknowledge all participants and São Leopoldo Mandic College.
CONFLICT OF INTEREST
The authors declare that there is no conflict of interest.
ABBREVIATIONS
Mm: Millimeter; RPM: Revolution per minute.
REFERENCES
1. Galindo DF, Butura CC. Immediately Loaded Mandibular Fixed Implant
Prostheses using the All-On-Four Protocol: A Report of 183 Consecutively Treated Patients with 1 Year of Function in Definitive Prosthes. International Journal of Oral and Maxillofacial Implants. 2012;27(3):628-33.
2. Lioubavina‐Hack N, Lang NP, Karring T. Significance of primary stability for osseointegration of dental implants. Clin Oral Implants Res. 2006;17(3):244-50.
3. Faverani LP, Ferreira GR, Jardim ECG, Okamoto R, Shinohara EH, Assunção WG, et al. Implantes osseointegrados: Evolução sucesso. Salusvita. 2011;30(1):47-58.
4. Babbush C, Kanwati A, Brokloff J. A New Approach to the AII-on-Four Treatment Concept Using Narrow Platform Nobel Active Implants. J Oral Implantol. 2013;39(3):314-25.
5. Bulutsuz AG, Tanyel RC, Katiboglu AB. Measurement of temperature change during the implant site preparation to determine influence of tool characteristics. Measurement. 2016;79:354-9.
6. Geerts GAVM, Patel Z. Temperature changes along a dental implant. Int J
Prosthodont. 2011;24(1):58-68.
7. Lucas RRS, Gonçalves R, Pinheiro MPF, Pinheiro AR, Alto RVM. Fatores que afetam a osseointegração dos implantes-Uma Revisão. Rev Flum Odontol. 2013;19(39):3-10.
8. Nero ALD, Gehrke SA, Bortolo JN, Zanatta LC. Temperatura durante a fresagem óssea. Estudo comparativo das técnicas de irrigação. Rev Assoc Paul Cir Dent.2012;66(2):147-50.
9. Instituto Nacional de Propriedade Industrial. Depósito de pedido nacional depatente. 2017. BR2020170107359.
10. Alves LMN, Hidalgo LRC, Conceição LS, Oliveira GM, Borges KRF, Passos WG. Implants complications: A literature review. J Orofac Invest. 2017;4(1):20-9.
11. Andriani JW. Mensuração do calor friccional gerado pelo preparo do tecido ósseo em implantodontia: estudo in vitro. Dissertação de Mestrado, Programa de Pós-Graduação em Odontologia da Universidade Federal de Santa Catarina: Florianópolis. 2002.
12. Carneiro MB, Medeiros RLR, Machado ÁR, Gomes VL, Soares MAD, Luiz NE. Influência das condições de corte no preparo de lojas receptoras de implantes dentais. Desenvolver habilidades e competências é imprescindível para sua carreira. Innov Implant J Biomater Esthet. 2009;4(1):13-8.
13. Faria R. Avaliação da temperatura gerada e do tempo gasto durante o preparo do tecido ósseo com a utilização de brocas de diferentes sistemas de implantes. Dissertação de Mestrado, Programa de Pós-Graduação em Odontologia Restauradora da Universidade Estadual Paulista. São José dos Campos. 2005.
14. Gehrke SA, Bettach R, Taschieri S, Boukhris G, Corbella S, DelFabbro M.
Temperature Changes in Cortical Bone after Implant Site Preparation Using a Single Bur versus Multiple Drilling Steps: An in vitro Investigation. Clin Implant Dent R. 2015;17(4):700-7.
15. Soldatos N, Gozalo D, Moreno D, Powell C. Temperature Changes During
Implant Osteotomies Utlilizing Three Different Implant Systems: A Pilot Study. JIACD. 2016;8(7):34-43.
16. Strbac GD, Giannis K, Unger E, Mittlböck M, Vasak C, Watzek G, et al. Drilling‐and withdrawing‐related thermal changes during implant site osteotomies. Clin Implant Dent R. 2015;17(1):32-43.
17. Trisi P, Berardini M, Falco A, Vulpiani MP. Effect of temperature on the dental implant osseointegration development in low-density bone: An in vivo histological evaluation. Implant Dent. 2015;24(1):96-100.
18. Fernandes MGA, Fonseca EMM, Natal R. Modelo 3D para análise térmica
durante o processo de furação do osso cortical. Rev Ass Port Anál Exp de Tens. 2015;(25):11-9.

Article History: Submission Date : 22-01-2019; Revised Date : 17-02-2019; Acceptance Date : 13-04-2019.
Cite this article: Subelza PH, Kopp G, Sivila MFH, Francischone CE. Comparative Study of Conventional Drill Bits and a New Model for Low Rotation in the Surgical Bed Preparation in Bone Blocks for Installation of Dental Implants. J Young Pharm. 2019;11(4):429-33.

Access and download the original article: http://www.jyoungpharm.org/article/1365

Related articles

18/04/2022 General

Discover the advantages of the Cone Morse Implant

For a long time, External Hexagon implants were unanimous in dental clinics across the country. However, this hegemony of years begins to be threatened by the Cone Morse implants, which have been increasingly used by implant dentists, given the numerous advantages that this implant system presents. Since there are still many doubts about Cone Morse […]

Kopp Implantes photo
by Kopp Implantes

Comments

Leave a Reply

Your email address will not be published. Required fields are marked *

Picture

Kopp Educação

Participamos e fomentamos a educação e formação de novos dentistas brasileiros, doando material para faculdades, onde todos podem conhecer nossos produtos e sistemas de implantes.

Conheça nossos produtos!

World icon Select your language

Contact us via WhatsApp WhatsApp icon